Effect of Different Anti-Coagulants on the Accuracy of Glycated Haemoglobin Results

Main Article Content

Frederick Igila Allison
Duru Ugochukwu Stephen

Abstract

Background: Most manufacturers of glycated haemoglobin kits advocate for the use of EDTA bottles for sample collection. Other manufacturers even when using the same glycated haemoglobin assay method, advocate for the use of any of these anticoagulant: EDTA, heparin and fluoride oxalate as any of these anticoagulants for sample collection.

Aim: This study was therefore designed to evaluate the effect of different anticoagulants on the accuracy of glycated haemoglobin value using the same method.

Methods: Thirty subjects were selected by purposive sampling method and 2ml of blood was collected from each subject into sodium heparin, EDTA and fluoride oxalate bottles and stored for three days at 4ºC. Fifteen subjects’ samples were analysed daily for the next two days then all the samples were analysed on the third day. All samples were analysed using the boronate affinity chromatographic method by Clover.

Results: The mean of the values of glycated haemoglobin of samples for each anticoagulants were about the same for the first, second and third day. The differences in the mean values for each anticoagulant were not statistically significant, indicating fairly good stability.

Conclusion: From this study, it could be concluded that blood sample in EDTA, fluoride oxalate and heparin bottles can be used for glycated haemoglobin estimation without affecting the accuracy of the result. These samples in these containers were found to be stable for at least three days.

Keywords:
Anti coagulants, glycated haemoglobin, chromatography, pre analytical stability.

Article Details

How to Cite
Allison, F. I., & Stephen, D. U. (2019). Effect of Different Anti-Coagulants on the Accuracy of Glycated Haemoglobin Results. Asian Journal of Medicine and Health, 16(4), 1-5. https://doi.org/10.9734/ajmah/2019/v16i430152
Section
Original Research Article

References

Diagnosis and classification of diabetes mellitus; Diabetes Care. 2010;33(1):62- 69.

Sarwar N, Gao S, Seshasai SR, Gobin R. Kaptoge S, Di Angelantonio E, Ingelsson E, lawlor DA, Selvin E, Stampfer M, Stehouwer CD, Lewington S, Pennells L, Thomson A, Sattar N, White IR, Ray KK, Danesh J. Diabetes Mellitus, fasting blood glucose concentration, and risk of vascular disease: a collaborative meta-analysis of 102 prospective studies. Lancet. 2010; 375(9733):2215-2222.

Singh R, Barden A, Mori T, Beillin L. Advanced glycation end-products; A review. Diabetologia. 2001;45:129-146.

Ginde AA, Cagliero E, Nathan DM, Camargo CA. Point of care glucose and haemoglobin A1c in emergency department patients without known diabetes: Implicatin of opportunistic screening. Acad Emerg Med. 2008;15: 1241-1247.

Lakshmy R, Gupta R. Measurement of glycated hemoglobin A1c from dried blood by turbidimetric immunoassay. J Diabetes sci technol. 2009;3(5):1203-1206.

Little RR, Rohlfing CL, Tennill AL, Connolly S, Hanson S. Effects of sample storage conditions on glycated hemoglobin measurement: Evaluation of five different high performance liquid chromatography methods. Diabetes Technol Ther. 2007;9: 36–42.

Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2002;48:436–472.

Adeoye S, Sherly A, Irivia E, Sylvester S, Borda T. Anaemia and glycated haemoglobin level; Is there a case for redefining reference ranges and therapeutic goals. BJMP. 2014;7:706- 711.

Bowen RA, Hortin GL, Csako G, otanez OH, Remaley AT. Impact of blood collection devices on clinical chemistry assays. Clin Biochem. 2010;43(1-2):4-25.

Geling L, Cabanero M, Wang Z, Wang H, Huang T, Herol A, Eid I, Muth G, Pinus MR. Comparison of glucose determinations on blood samples collected in three types of tubes. Ann Clin Lab Sci. 2013;43(3):278-284.

Snehankar K, Rumi D, Risha G, Kailash B. Sodium fluoride vacutainer- an elternative to EDTA for estimation of glycated haemoglobin. Int J Sci Res. 2017;6:127-129.

Vrtaric A, Filipi P, Hemar M, Nikolac N, Simundic AM. K2-EDTA and K3-EDTA Greiner Tubes for HbA1c Measurement. Lab Med. 2016;47(1):39-42.

Little RR, Rohlfing CL, Tennill AL, Connolly S, Hanson S. Effects of sample storage conditions on glycated hemoglobin measurement: Evaluation of five different high performance liquid chromatography methods. Diabetes Technol Ther. 2007; 9:36 42.5.